BlueStar Diabetes
Making two-hour glucose predictions easy to understand, and safe to act on.
BlueStar's new AI could predict glucose levels two hours ahead: a value and a range every five minutes. My job was to make that easy to understand at a glance, without nudging anyone toward an unsafe decision. I borrowed from the weather forecast: a timeline you can play and pause. After redesigning from usability testing, nobody misread the range, and everyone found the feature on their own.
- Role
- Product Designer, Welldoc
- Teams
- Clinical, engineering, human factors, data science, PMs
- Timeline
- Aug 2024 - Sep 2024 • Active designSep 2025 – Oct 2025 • Build & testing
- Impact
- Enabled the FDA submission of a new, in-demand feature.

IPrediction mode
A two-hour prediction should read like a weather forecast.

Turning 24 predictions into a forecast you can play
When our Chief Medical Officer introduced the new technology, my first thought was that it was a lot: a predicted value and a range every five minutes, for two hours. But we already know how to read something like that. A weather forecast turns uncertain data into a timeline you can glance at.
So prediction mode became a timeline you can play and pause. A 10-second loop runs from "Now" to two hours ahead, and the glucose value and graph update as it plays. It keeps the structure of the existing glucose screen (the value on top, the graph below), so it feels familiar from the first tap.
The first usability test was encouraging.
of participants said the display was "intuitive"
liked seeing the predicted value in the top panel, not just on the graph


Designing within real constraints
- The graph had to stay within the existing development library, so instead of a new chart I added a purple band for the predicted range and a blue line for the predicted time.
- Scrubbing the timeline wasn't supported, so I removed the scrub handle. The UI should never promise something it can't do.
- Predictions refresh every five minutes, so I worked with engineering on how the timeline moves over time and when to prompt people to refresh an outdated prediction.
IIThe range
The range mattered more than the number, so it appears the moment you pause.


Value while playing, range when paused
Every prediction has two parts: a predicted value and a range. During playback each prediction is on screen for less than a second, so the value leads. But the clinical team pointed out that the range is what's actually meaningful to people. So when you pause, the range appears.


Participants read the range as limits for the whole graph
In usability testing, 43% of participants read the high and low numbers as limits for the entire graph, not just for the moment they'd paused on. For a feature people use to make health decisions, that was a safety problem.
I replaced the stacked "High End" and "Low End" labels with a simple "Between 224 – 264", right under the predicted value, so the range clearly belongs to that moment. We also tried showing only the range, but it barely changes from point to point, and the animation lost its meaning.
What changed
of participants misread the range, before and after the redesign
IIISafety
Every prediction came with advice, and the advice had to be safe for the person reading it.


Room for the right message
The clinical team defined 12 situations by combining four current glucose states (very low, low, in range, high) with three predicted ones (low, in range, high), each with its own coaching message. A pop-up couldn't hold that, so over several iterations prediction mode moved into a bottom sheet with room for the full message.
The test that uncovered the biggest risk
I built the human factors prototypes with precise numbers, graph behavior and animation, so participants reacted to something that felt real. That's how we found the biggest risk: many participants opened prediction mode while using insulin, and the feature isn't designed for insulin users.
IVFollowing through
When safety and fewer taps disagreed, safety won.
I built the prototypes for every round of human factors testing and designed the changes. Before the redesign stage for the final rounds, I was pulled onto a higher-priority initiative, Lilly Health, so my teammate made the changes below. I kept following the project through to the FDA submission.


A clearer way in
I had kept the play button close to the existing glucose screen so the new mode would feel native. In testing, that backfired: 40% of participants needed a prompt to find it.
The new entry point is a larger, self-contained button in the brand's primary blue, labeled "View My Predicted Glucose" instead of "Predict Glucose Levels". It says what you'll see, not what the system does.
What changed
of participants needed a prompt to find the feature, before and after the redesign
Safeguards for insulin users
The entry point is hidden for anyone with insulin in their medication list. Everyone else is asked first whether they use any type of insulin. If they answer yes, they're asked to add their insulin, and once they do, the entry point disappears. If they answer no, a warning says the feature isn't designed for insulin users, and prediction mode opens only after they tap I Understand.
The disclaimers also say plainly not to adjust medication based on a prediction. The flow adds a few taps, and that was the right trade.
It follows the order human factors asks for: remove the risk by design first (hiding the entry point), then guard against it (the question), and keep the warning visible (the disclaimers).



VOutcome
The designs got a new, in-demand BlueStar feature ready for its FDA submission.
Final screens





What I took away
Prototyping for human factors testing took longer than designing for hand-off, and it was worth it. The numbers, the graph and the animation had to be exactly right for people with diabetes to react honestly, and their reactions were critical to gauge safety.
This is the project where my human factors background felt most useful. Working closely with that team was the highlight for me.